Dosage for Cardiac Dual-Source CT Can Be Significantly Lowered
|
By MedImaging International staff writers Posted on 13 Aug 2008 |
A new study reveals that by using dual-source computed tomography (DSCT), the effective dosage for a heart examination can be significantly lowered, in comparison to conventional CT. The study also demonstrated that stenoses could be diagnosed with similar high accuracy as with invasive X-ray angiography.
At the University Hospital Zürich (Switzerland), 120 patients with suspected coronary heart disease were scanned with the world's first CT scanner with two X-ray tubes, a Somatom Definition system, developed by Siemens Healthcare (Erlangen, Germany). The Siemens application, Adaptive Cardio Sequence with a step-and-shoot mode, was also used for the first time with a dual-source CT. The results of the study were published in the June 2008, issue of the journal Heart.
Angiography is a diagnostic procedure used to determine coronary arterial occlusion. Typically, this examination is conducted using angiography devices in the catheter laboratory and CT images. However, with CT, the question of the radiation dosage plays a huge role. Researchers at the University Hospital now want to see whether the dose used during CT angiography can be reduced.
The research team used Adaptive Cardio Sequence, which was recently developed by Siemens for all Somatom Definition Scanners. The application is based on the step-and-shoot method, in which the next respective diastole is calculated, and only this phase of the cardiac cycle is used to obtain image data. The combination of Adaptive Cardio Sequence and the DSCT's maximum time resolution of 83 ms facilitate this technique in a particularly effective manner and permits a further significant reduction in dosage.
"The results show that CT coronary angiography with a dual-source CT in step-and-shoot mode produces images of excellent diagnostic quality in patients with stable heart rates up to 70 bpm. The dosage reduction achieved in comparison to previous CT angiography is also remarkable. In our study, we required an effective dose of 2.5 mSv on average with a deviation of plus or minus 0.8 mSv. In the literature, a normal average effective dose for heart scans of between 9 mSv and 21 mSv is reported,” said Dr. Hatem Alkadhi, specialist in radiology at the Institute for Diagnostic Radiology, University Hospital Zürich.
The Adaptive Cardio Sequence supplements the step-and-shoot method with intelligent algorithms, which monitor the heart frequency of patients, and if necessary, respond to arrhythmia, such as extra systoles (i.e., additional heart beats). In this case, the recording phase is automatically postponed to avoid image errors due to the sudden movement. Moreover, the recording window can be widened a little more if required so that, using the Adaptive Cardio Sequence, the robustness of the CT scan can be increased significantly overall.
"For Siemens, the safety of the patients has always been the primary focus of the developments in CT. At the same time, we consider our highest obligation to be to give doctors the best diagnostic image quality and to support them in making their workflow as efficient as possible. With this outlook and constant further development, Siemens has positioned itself as a market leader in CT technology,” said Dr. Sami Atiya, CEO, Computed Tomography, Siemens Healthcare. "Innovations for dose reduction, in conjunction with improved diagnostic image quality, are therefore of the utmost importance to us. This is why we have also further developed our Adaptive Cardio Sequence, now for the dual-source CT. The study from Zurich confirms to us that the combination of step-and-shoot and dual-source computed tomography does significantly reduce the dosage required during heart scans.”
Related Links:
University Hospital Zurich
Siemens Healthcare
At the University Hospital Zürich (Switzerland), 120 patients with suspected coronary heart disease were scanned with the world's first CT scanner with two X-ray tubes, a Somatom Definition system, developed by Siemens Healthcare (Erlangen, Germany). The Siemens application, Adaptive Cardio Sequence with a step-and-shoot mode, was also used for the first time with a dual-source CT. The results of the study were published in the June 2008, issue of the journal Heart.
Angiography is a diagnostic procedure used to determine coronary arterial occlusion. Typically, this examination is conducted using angiography devices in the catheter laboratory and CT images. However, with CT, the question of the radiation dosage plays a huge role. Researchers at the University Hospital now want to see whether the dose used during CT angiography can be reduced.
The research team used Adaptive Cardio Sequence, which was recently developed by Siemens for all Somatom Definition Scanners. The application is based on the step-and-shoot method, in which the next respective diastole is calculated, and only this phase of the cardiac cycle is used to obtain image data. The combination of Adaptive Cardio Sequence and the DSCT's maximum time resolution of 83 ms facilitate this technique in a particularly effective manner and permits a further significant reduction in dosage.
"The results show that CT coronary angiography with a dual-source CT in step-and-shoot mode produces images of excellent diagnostic quality in patients with stable heart rates up to 70 bpm. The dosage reduction achieved in comparison to previous CT angiography is also remarkable. In our study, we required an effective dose of 2.5 mSv on average with a deviation of plus or minus 0.8 mSv. In the literature, a normal average effective dose for heart scans of between 9 mSv and 21 mSv is reported,” said Dr. Hatem Alkadhi, specialist in radiology at the Institute for Diagnostic Radiology, University Hospital Zürich.
The Adaptive Cardio Sequence supplements the step-and-shoot method with intelligent algorithms, which monitor the heart frequency of patients, and if necessary, respond to arrhythmia, such as extra systoles (i.e., additional heart beats). In this case, the recording phase is automatically postponed to avoid image errors due to the sudden movement. Moreover, the recording window can be widened a little more if required so that, using the Adaptive Cardio Sequence, the robustness of the CT scan can be increased significantly overall.
"For Siemens, the safety of the patients has always been the primary focus of the developments in CT. At the same time, we consider our highest obligation to be to give doctors the best diagnostic image quality and to support them in making their workflow as efficient as possible. With this outlook and constant further development, Siemens has positioned itself as a market leader in CT technology,” said Dr. Sami Atiya, CEO, Computed Tomography, Siemens Healthcare. "Innovations for dose reduction, in conjunction with improved diagnostic image quality, are therefore of the utmost importance to us. This is why we have also further developed our Adaptive Cardio Sequence, now for the dual-source CT. The study from Zurich confirms to us that the combination of step-and-shoot and dual-source computed tomography does significantly reduce the dosage required during heart scans.”
Related Links:
University Hospital Zurich
Siemens Healthcare
Latest Radiography News
- AI Generates Future Knee X-Rays to Predict Osteoarthritis Progression Risk
- AI Algorithm Uses Mammograms to Accurately Predict Cardiovascular Risk in Women
- AI Hybrid Strategy Improves Mammogram Interpretation
- AI Technology Predicts Personalized Five-Year Risk of Developing Breast Cancer
- RSNA AI Challenge Models Can Independently Interpret Mammograms
- New Technique Combines X-Ray Imaging and Radar for Safer Cancer Diagnosis
- New AI Tool Helps Doctors Read Chest X‑Rays Better
- Wearable X-Ray Imaging Detecting Fabric to Provide On-The-Go Diagnostic Scanning
- AI Helps Radiologists Spot More Lesions in Mammograms
- AI Detects Fatty Liver Disease from Chest X-Rays
- AI Detects Hidden Heart Disease in Existing CT Chest Scans
- Ultra-Lightweight AI Model Runs Without GPU to Break Barriers in Lung Cancer Diagnosis
- AI Radiology Tool Identifies Life-Threatening Conditions in Milliseconds

- Machine Learning Algorithm Identifies Cardiovascular Risk from Routine Bone Density Scans
- AI Improves Early Detection of Interval Breast Cancers
- World's Largest Class Single Crystal Diamond Radiation Detector Opens New Possibilities for Diagnostic Imaging
Channels
MRI
view channel
AI-Assisted Model Enhances MRI Heart Scans
A cardiac MRI can reveal critical information about the heart’s function and any abnormalities, but traditional scans take 30 to 90 minutes and often suffer from poor image quality due to patient movement.... Read more
AI Model Outperforms Doctors at Identifying Patients Most At-Risk of Cardiac Arrest
Hypertrophic cardiomyopathy is one of the most common inherited heart conditions and a leading cause of sudden cardiac death in young individuals and athletes. While many patients live normal lives, some... Read moreUltrasound
view channel
Ultrasound Probe Images Entire Organ in 4D
Disorders of blood microcirculation can have devastating effects, contributing to heart failure, kidney failure, and chronic diseases. However, existing imaging technologies cannot visualize the full network... Read more
Disposable Ultrasound Patch Performs Better Than Existing Devices
Wearable ultrasound devices are widely used in diagnostics, rehabilitation monitoring, and telemedicine, yet most existing models rely on lead-based piezoelectric ceramics that pose health and environmental risks.... Read moreNuclear Medicine
view channel
New Imaging Solution Improves Survival for Patients with Recurring Prostate Cancer
Detecting recurrent prostate cancer remains one of the most difficult challenges in oncology, as standard imaging methods such as bone scans and CT scans often fail to accurately locate small or early-stage tumors.... Read more
PET Tracer Enables Same-Day Imaging of Triple-Negative Breast and Urothelial Cancers
Triple-negative breast cancer (TNBC) and urothelial bladder carcinoma (UBC) are aggressive cancers often diagnosed at advanced stages, leaving limited time for effective treatment decisions.... Read more
New Camera Sees Inside Human Body for Enhanced Scanning and Diagnosis
Nuclear medicine scans like single-photon emission computed tomography (SPECT) allow doctors to observe heart function, track blood flow, and detect hidden diseases. However, current detectors are either... Read more
Novel Bacteria-Specific PET Imaging Approach Detects Hard-To-Diagnose Lung Infections
Mycobacteroides abscessus is a rapidly growing mycobacteria that primarily affects immunocompromised patients and those with underlying lung diseases, such as cystic fibrosis or chronic obstructive pulmonary... Read moreGeneral/Advanced Imaging
view channel
New Ultrasmall, Light-Sensitive Nanoparticles Could Serve as Contrast Agents
Medical imaging technologies face ongoing challenges in capturing accurate, detailed views of internal processes, especially in conditions like cancer, where tracking disease development and treatment... Read more
AI Algorithm Accurately Predicts Pancreatic Cancer Metastasis Using Routine CT Images
In pancreatic cancer, detecting whether the disease has spread to other organs is critical for determining whether surgery is appropriate. If metastasis is present, surgery is not recommended, yet current... Read moreImaging IT
view channel
New Google Cloud Medical Imaging Suite Makes Imaging Healthcare Data More Accessible
Medical imaging is a critical tool used to diagnose patients, and there are billions of medical images scanned globally each year. Imaging data accounts for about 90% of all healthcare data1 and, until... Read more
Global AI in Medical Diagnostics Market to Be Driven by Demand for Image Recognition in Radiology
The global artificial intelligence (AI) in medical diagnostics market is expanding with early disease detection being one of its key applications and image recognition becoming a compelling consumer proposition... Read moreIndustry News
view channel
GE HealthCare and NVIDIA Collaboration to Reimagine Diagnostic Imaging
GE HealthCare (Chicago, IL, USA) has entered into a collaboration with NVIDIA (Santa Clara, CA, USA), expanding the existing relationship between the two companies to focus on pioneering innovation in... Read morePatient-Specific 3D-Printed Phantoms Transform CT Imaging
New research has highlighted how anatomically precise, patient-specific 3D-printed phantoms are proving to be scalable, cost-effective, and efficient tools in the development of new CT scan algorithms... Read more
Siemens and Sectra Collaborate on Enhancing Radiology Workflows
Siemens Healthineers (Forchheim, Germany) and Sectra (Linköping, Sweden) have entered into a collaboration aimed at enhancing radiologists' diagnostic capabilities and, in turn, improving patient care... Read more