Dosage for Cardiac Dual-Source CT Can Be Significantly Lowered
|
By MedImaging International staff writers Posted on 13 Aug 2008 |
A new study reveals that by using dual-source computed tomography (DSCT), the effective dosage for a heart examination can be significantly lowered, in comparison to conventional CT. The study also demonstrated that stenoses could be diagnosed with similar high accuracy as with invasive X-ray angiography.
At the University Hospital Zürich (Switzerland), 120 patients with suspected coronary heart disease were scanned with the world's first CT scanner with two X-ray tubes, a Somatom Definition system, developed by Siemens Healthcare (Erlangen, Germany). The Siemens application, Adaptive Cardio Sequence with a step-and-shoot mode, was also used for the first time with a dual-source CT. The results of the study were published in the June 2008, issue of the journal Heart.
Angiography is a diagnostic procedure used to determine coronary arterial occlusion. Typically, this examination is conducted using angiography devices in the catheter laboratory and CT images. However, with CT, the question of the radiation dosage plays a huge role. Researchers at the University Hospital now want to see whether the dose used during CT angiography can be reduced.
The research team used Adaptive Cardio Sequence, which was recently developed by Siemens for all Somatom Definition Scanners. The application is based on the step-and-shoot method, in which the next respective diastole is calculated, and only this phase of the cardiac cycle is used to obtain image data. The combination of Adaptive Cardio Sequence and the DSCT's maximum time resolution of 83 ms facilitate this technique in a particularly effective manner and permits a further significant reduction in dosage.
"The results show that CT coronary angiography with a dual-source CT in step-and-shoot mode produces images of excellent diagnostic quality in patients with stable heart rates up to 70 bpm. The dosage reduction achieved in comparison to previous CT angiography is also remarkable. In our study, we required an effective dose of 2.5 mSv on average with a deviation of plus or minus 0.8 mSv. In the literature, a normal average effective dose for heart scans of between 9 mSv and 21 mSv is reported,” said Dr. Hatem Alkadhi, specialist in radiology at the Institute for Diagnostic Radiology, University Hospital Zürich.
The Adaptive Cardio Sequence supplements the step-and-shoot method with intelligent algorithms, which monitor the heart frequency of patients, and if necessary, respond to arrhythmia, such as extra systoles (i.e., additional heart beats). In this case, the recording phase is automatically postponed to avoid image errors due to the sudden movement. Moreover, the recording window can be widened a little more if required so that, using the Adaptive Cardio Sequence, the robustness of the CT scan can be increased significantly overall.
"For Siemens, the safety of the patients has always been the primary focus of the developments in CT. At the same time, we consider our highest obligation to be to give doctors the best diagnostic image quality and to support them in making their workflow as efficient as possible. With this outlook and constant further development, Siemens has positioned itself as a market leader in CT technology,” said Dr. Sami Atiya, CEO, Computed Tomography, Siemens Healthcare. "Innovations for dose reduction, in conjunction with improved diagnostic image quality, are therefore of the utmost importance to us. This is why we have also further developed our Adaptive Cardio Sequence, now for the dual-source CT. The study from Zurich confirms to us that the combination of step-and-shoot and dual-source computed tomography does significantly reduce the dosage required during heart scans.”
Related Links:
University Hospital Zurich
Siemens Healthcare
At the University Hospital Zürich (Switzerland), 120 patients with suspected coronary heart disease were scanned with the world's first CT scanner with two X-ray tubes, a Somatom Definition system, developed by Siemens Healthcare (Erlangen, Germany). The Siemens application, Adaptive Cardio Sequence with a step-and-shoot mode, was also used for the first time with a dual-source CT. The results of the study were published in the June 2008, issue of the journal Heart.
Angiography is a diagnostic procedure used to determine coronary arterial occlusion. Typically, this examination is conducted using angiography devices in the catheter laboratory and CT images. However, with CT, the question of the radiation dosage plays a huge role. Researchers at the University Hospital now want to see whether the dose used during CT angiography can be reduced.
The research team used Adaptive Cardio Sequence, which was recently developed by Siemens for all Somatom Definition Scanners. The application is based on the step-and-shoot method, in which the next respective diastole is calculated, and only this phase of the cardiac cycle is used to obtain image data. The combination of Adaptive Cardio Sequence and the DSCT's maximum time resolution of 83 ms facilitate this technique in a particularly effective manner and permits a further significant reduction in dosage.
"The results show that CT coronary angiography with a dual-source CT in step-and-shoot mode produces images of excellent diagnostic quality in patients with stable heart rates up to 70 bpm. The dosage reduction achieved in comparison to previous CT angiography is also remarkable. In our study, we required an effective dose of 2.5 mSv on average with a deviation of plus or minus 0.8 mSv. In the literature, a normal average effective dose for heart scans of between 9 mSv and 21 mSv is reported,” said Dr. Hatem Alkadhi, specialist in radiology at the Institute for Diagnostic Radiology, University Hospital Zürich.
The Adaptive Cardio Sequence supplements the step-and-shoot method with intelligent algorithms, which monitor the heart frequency of patients, and if necessary, respond to arrhythmia, such as extra systoles (i.e., additional heart beats). In this case, the recording phase is automatically postponed to avoid image errors due to the sudden movement. Moreover, the recording window can be widened a little more if required so that, using the Adaptive Cardio Sequence, the robustness of the CT scan can be increased significantly overall.
"For Siemens, the safety of the patients has always been the primary focus of the developments in CT. At the same time, we consider our highest obligation to be to give doctors the best diagnostic image quality and to support them in making their workflow as efficient as possible. With this outlook and constant further development, Siemens has positioned itself as a market leader in CT technology,” said Dr. Sami Atiya, CEO, Computed Tomography, Siemens Healthcare. "Innovations for dose reduction, in conjunction with improved diagnostic image quality, are therefore of the utmost importance to us. This is why we have also further developed our Adaptive Cardio Sequence, now for the dual-source CT. The study from Zurich confirms to us that the combination of step-and-shoot and dual-source computed tomography does significantly reduce the dosage required during heart scans.”
Related Links:
University Hospital Zurich
Siemens Healthcare
Latest Radiography News
- AI Mammography Tools Detect Early Breast Cancer Signs Years Before Diagnosis
- Rapid X-Ray Test Quantifies Pulmonary Regurgitation After Tetralogy of Fallot Repair
- AI Tool Flags Osteoporosis Risk from Routine Chest X-Rays
- Simple Chest X-Ray Measure Predicts Survival After Lung Cancer Surgery
- AI Detection Tool Improves Identification of Lobular Breast Cancer
- New Contrast Agent Enables Low-Dose X-Ray Joint Imaging
- AI Boosts Breast Cancer Detection and Cuts Screening Workload
- AI Tool Predicts Breast Cancer Risk Years Ahead Using Routine Mammograms
- Routine Mammograms Could Predict Future Cardiovascular Disease in Women
- AI Detects Early Signs of Aging from Chest X-Rays
- X-Ray Breakthrough Captures Three Image-Contrast Types in Single Shot
- AI Generates Future Knee X-Rays to Predict Osteoarthritis Progression Risk
- AI Algorithm Uses Mammograms to Accurately Predict Cardiovascular Risk in Women
- AI Hybrid Strategy Improves Mammogram Interpretation
- AI Technology Predicts Personalized Five-Year Risk of Developing Breast Cancer
- RSNA AI Challenge Models Can Independently Interpret Mammograms
Channels
MRI
view channel
International Study Assesses AI for Prostate Cancer MRI Interpretation
Prostate cancer is a leading cause of cancer morbidity in men, and accurate early diagnosis hinges on expert interpretation of prostate magnetic resonance imaging (MRI). Rapid adoption of MRI-first pathways... Read more
AI Approach Could Shorten Advanced Brain MRI Scans by Up to 90%
Long acquisition times for advanced brain magnetic resonance imaging (MRI) can limit access, extend waiting lists, and disrupt clinical workflows. Reducing data requirements without sacrificing image fidelity... Read moreUltrasound
view channelAI Robotic Ultrasound System Automates Echocardiography and Improves Consistency
Echocardiography, an ultrasound examination of the heart, is central to diagnosing and managing cardiovascular disease. Many services struggle with limited availability of skilled sonographers, variable... Read more
Whole Cross-Section Ultrasound System Enables Operator-Independent Imaging
Conventional ultrasound is central to bedside imaging but is limited by a narrow field of view and operator variability. Comprehensive cross-sectional assessment typically requires computed tomography... Read moreNuclear Medicine
view channelNew PET Tracer Detects DVT and Pulmonary Embolism in One Scan
Deep vein thrombosis is the formation of clots in deep leg veins that can migrate to the lungs as pulmonary embolism. Rapid confirmation across both regions often requires multiple tests and can delay treatment.... Read more
Targeted PET Platform Guides Osteosarcoma Resection and Margin Verification
Osteosarcoma, an aggressive primary bone cancer that mainly affects children and adolescents, demands wide excision to prevent local recurrence. Surgeons must achieve negative margins while preserving... Read moreGeneral/Advanced Imaging
view channel
CT-Derived Biomarker Predicts Outcomes in Gastric Cancer
Gastric cancer, also known as stomach cancer, is the fifth most common malignancy worldwide and often shows heterogeneous outcomes even within the same stage. Prognostic estimates typically rely on tumor-centric... Read more
AI Tool Enhances Response Assessment and Survival Prediction in Pleural Mesothelioma
Pleural mesothelioma, a cancer that grows as a thin, irregular layer along the lung wall, is difficult to measure on imaging. Clinicians rely on diameter-based Response Evaluation Criteria in Solid Tumors... Read more
AI Tool Enables Real-Time Diffuse Optical Tomography for Brain Lesion Detection
Diffuse optical tomography is a noninvasive imaging technique that uses near-infrared light to detect internal abnormalities such as cerebral hemorrhage and tumors. Its clinical utility for real-time ... Read moreNew SPECT/CT Method Differentiates Inflammation from Fibrosis in Interstitial Lung Disease
Interstitial lung disease (ILD) encompasses more than 200 disorders that inflame or scar the lung interstitium and can lead to progressive respiratory failure. Determining whether active inflammation is... Read moreImaging IT
view channel
Ambient AI Reporting Platform Streamlines Radiology Reporting
Radiology departments face growing imaging volumes and staffing shortages, creating reporting bottlenecks and pressure to maintain turnaround times. Conventional dictation tools document findings after... Read more
Interactive AI Tool Supports Explainable Lung Nodule Assessment
Lung cancer is a leading cause of cancer mortality, and timely characterization of pulmonary nodules on chest computed tomography (CT) is essential for directing care. Interpreting nodule morphology demands... Read more
Breast Imaging Software Enhances Visualization and Tissue Characterization in Challenging Cases
Breast imaging can be particularly challenging in cases involving small breasts or implants, where image reconstruction and tissue characterization may be limited. Clinicians also need reproducible analysis... Read moreIndustry News
view channel
GE HealthCare Showcases AI-Enabled Nuclear Medicine Portfolio at SNMMI 2026
Nuclear medicine is expanding rapidly as health systems adopt theranostics and broaden access to radiopharmaceuticals, increasing demand for scalable operations and consistent diagnostic confidence.... Read more
GE HealthCare Highlights AI-Supported Radiation Therapy Tools at ESTRO 2026
At the European Society for Radiotherapy and Oncology (ESTRO) 2026 Congress in Stockholm, GE HealthCare is highlighting Intelligent Radiation Therapy (iRT), MIM Software innovations, and BK Medical surgical... Read more





 Guided Devices.jpg)