MRI Safety Risk for Patients with Pacemakers Identified
|
By MedImaging International staff writers Posted on 19 Jan 2010 |
U.S. researchers have found that certain cardiac pacemakers may inadequately stimulate a patient's heart while undergoing a magnetic resonance imaging (MRI) scan due to the magnetic pulses mixing with the electronic pulses from the pacemaker. This inadequate stimulation is potentially dangerous for the patient undergoing the MRI scan.
MRI is an imaging technique that utilizes a magnetic field instead of ionizing radiation to produce a detailed image of internal body structures. MRI systems expose patients to very strong magnetic fields that can interfere with implanted cardiac pacemakers. Physicians are instructed by pacemaker manufacturers and MRI system manufacturers not to expose patients with pacemakers to MRI scans. MRI can damage the pacemaker's electronic system and cause burning of heart tissue at the tip of the pacemaker lead, due to an increase in temperature from the MRI. Both risks can result in incorrect or absent stimulation from the pacemaker.
However, some cardiologists have published special protocols that describe how to allow patients with cardiac pacemakers to receive MRI scans. Some have stated that for specific patients, the diagnostic benefit from MR imaging versus other pacemaker compatible imaging modalities outweighs the risks.
U.S. Food and Drug Administration (FDA; Silver Spring, MD, USA) researchers Drs. Howard Bassen and Gonzalo Mendoza evaluated the risk of pacemakers causing unintended cardiac stimulation following exposure to a simulated MRI magnetic field by measuring electrical voltage produced at the tip of the pacemaker lead, where it would touch the interior of the heart.
Drs. Bassen and Mendoza found that when exposed to the strong, "gradient,” magnetic field, the pacemaker could deliver a drastically altered pulse and stimulate the heart inappropriately, which could have devastating consequences for the patient. "MRI systems emit several types of extremely intense magnetic fields and have caused injury to patients due to interactions with pacemakers,” said Dr. Bassen. "Cardiologists who choose to scan patients with cardiac pacemakers must assess the risks versus the benefits of the scan.”
The study's findings were published in December 2009 in BioMed Central's open access journal BioMedical Engineering Online.
Each year in the United States alone, patients undergo some 40 million MRI procedures. There are also 10 million pacemaker users in the United States.
Related Links:
U.S. Food and Drug Administration
MRI is an imaging technique that utilizes a magnetic field instead of ionizing radiation to produce a detailed image of internal body structures. MRI systems expose patients to very strong magnetic fields that can interfere with implanted cardiac pacemakers. Physicians are instructed by pacemaker manufacturers and MRI system manufacturers not to expose patients with pacemakers to MRI scans. MRI can damage the pacemaker's electronic system and cause burning of heart tissue at the tip of the pacemaker lead, due to an increase in temperature from the MRI. Both risks can result in incorrect or absent stimulation from the pacemaker.
However, some cardiologists have published special protocols that describe how to allow patients with cardiac pacemakers to receive MRI scans. Some have stated that for specific patients, the diagnostic benefit from MR imaging versus other pacemaker compatible imaging modalities outweighs the risks.
U.S. Food and Drug Administration (FDA; Silver Spring, MD, USA) researchers Drs. Howard Bassen and Gonzalo Mendoza evaluated the risk of pacemakers causing unintended cardiac stimulation following exposure to a simulated MRI magnetic field by measuring electrical voltage produced at the tip of the pacemaker lead, where it would touch the interior of the heart.
Drs. Bassen and Mendoza found that when exposed to the strong, "gradient,” magnetic field, the pacemaker could deliver a drastically altered pulse and stimulate the heart inappropriately, which could have devastating consequences for the patient. "MRI systems emit several types of extremely intense magnetic fields and have caused injury to patients due to interactions with pacemakers,” said Dr. Bassen. "Cardiologists who choose to scan patients with cardiac pacemakers must assess the risks versus the benefits of the scan.”
The study's findings were published in December 2009 in BioMed Central's open access journal BioMedical Engineering Online.
Each year in the United States alone, patients undergo some 40 million MRI procedures. There are also 10 million pacemaker users in the United States.
Related Links:
U.S. Food and Drug Administration
Latest MRI News
- AI Approach Could Shorten Advanced Brain MRI Scans by Up to 90%
- Cardiac MRI Measure Improves Risk Prediction in Tricuspid Regurgitation
- AI System Improves Accuracy of Cardiac MRI Interpretation
- Deep Learning Model Predicts Alzheimer’s Disease Outcomes from Baseline MRI
- Blood-Brain Barrier Imaging Adds Risk Insight to Standard Stroke MRI
- AI Body Composition MRI Analysis Predicts Cardiometabolic Disease Risk
- AI MRI Tool Quantifies Muscle Fat to Assess Cardiometabolic Risk
- Advanced MRI Visualizes CSF Motion Changes After Mild Traumatic Brain Injury
- MRI Tool Enables Long-Term Tracking of Transplanted Cardiac Cells
- MRI-Based AI Tool Supports Differentiation of Parkinsonian Syndromes
- MRI-Derived Biomarker Improves Risk Stratification in Glioblastoma
- Combined Imaging Approach Identifies Cause of Heart Attack without Coronary Blockage
- Advanced MRI System Detects Impaired Cardiac Oxygen Use in Minutes
- AI-Enhanced MRI Improves Image Quality in Arrhythmia Patients
- Ultra-Detailed Brain Atlas Enhances Early Detection of Neurological Disorders
- Study Finds Advanced Imaging Significantly Reduces Unnecessary Prostate Biopsies
Channels
Radiography
view channel
Rapid X-Ray Test Quantifies Pulmonary Regurgitation After Tetralogy of Fallot Repair
Tetralogy of Fallot is the most common cyanotic congenital heart defect and can leave patients with pulmonary valve regurgitation, a backward flow of blood into the right ventricle after repair.... Read more
AI Tool Flags Osteoporosis Risk from Routine Chest X-Rays
Osteoporosis is a progressive loss of bone density that is often silent until a fracture occurs. Current screening frameworks concentrate on older women and select high-risk groups. Many men, younger adults,... Read moreUltrasound
view channelAI Robotic Ultrasound System Automates Echocardiography and Improves Consistency
Echocardiography, an ultrasound examination of the heart, is central to diagnosing and managing cardiovascular disease. Many services struggle with limited availability of skilled sonographers, variable... Read more
Whole Cross-Section Ultrasound System Enables Operator-Independent Imaging
Conventional ultrasound is central to bedside imaging but is limited by a narrow field of view and operator variability. Comprehensive cross-sectional assessment typically requires computed tomography... Read moreNuclear Medicine
view channel
Targeted PET Platform Guides Osteosarcoma Resection and Margin Verification
Osteosarcoma, an aggressive primary bone cancer that mainly affects children and adolescents, demands wide excision to prevent local recurrence. Surgeons must achieve negative margins while preserving... Read more
Portable PET System Enables Real-Time Bedside Guidance for Biopsies and Ablations
Interventional radiology procedures typically rely on ultrasound, X-ray fluoroscopy, or computed tomography for image guidance. These modalities visualize anatomy but offer limited molecular information,... Read moreGeneral/Advanced Imaging
view channelNew SPECT/CT Method Differentiates Inflammation from Fibrosis in Interstitial Lung Disease
Interstitial lung disease (ILD) encompasses more than 200 disorders that inflame or scar the lung interstitium and can lead to progressive respiratory failure. Determining whether active inflammation is... Read more
Whole-Body PET/CT Tracks Metabolic Changes After Bariatric Surgery
Obesity surgery improves weight and comorbidity profiles, yet clinicians lack tools to monitor organ-level metabolic recovery after the procedure. A clear view of systemic changes could refine follow-up... Read moreImaging IT
view channel
Interactive AI Tool Supports Explainable Lung Nodule Assessment
Lung cancer is a leading cause of cancer mortality, and timely characterization of pulmonary nodules on chest computed tomography (CT) is essential for directing care. Interpreting nodule morphology demands... Read more
Breast Imaging Software Enhances Visualization and Tissue Characterization in Challenging Cases
Breast imaging can be particularly challenging in cases involving small breasts or implants, where image reconstruction and tissue characterization may be limited. Clinicians also need reproducible analysis... Read more
New Google Cloud Medical Imaging Suite Makes Imaging Healthcare Data More Accessible
Medical imaging is a critical tool used to diagnose patients, and there are billions of medical images scanned globally each year. Imaging data accounts for about 90% of all healthcare data1 and, until... Read more
Global AI in Medical Diagnostics Market to Be Driven by Demand for Image Recognition in Radiology
The global artificial intelligence (AI) in medical diagnostics market is expanding with early disease detection being one of its key applications and image recognition becoming a compelling consumer proposition... Read moreIndustry News
view channel
GE HealthCare Showcases AI-Enabled Nuclear Medicine Portfolio at SNMMI 2026
Nuclear medicine is expanding rapidly as health systems adopt theranostics and broaden access to radiopharmaceuticals, increasing demand for scalable operations and consistent diagnostic confidence.... Read more
GE HealthCare Highlights AI-Supported Radiation Therapy Tools at ESTRO 2026
At the European Society for Radiotherapy and Oncology (ESTRO) 2026 Congress in Stockholm, GE HealthCare is highlighting Intelligent Radiation Therapy (iRT), MIM Software innovations, and BK Medical surgical... Read more




 Guided Devices.jpg)