MedImaging

Download Mobile App
Recent News Radiography MRI Ultrasound Nuclear Medicine General/Advanced Imaging Imaging IT Industry News

Recall of Thousands of MRI Scanners Issued by US FDA

By MedImaging International staff writers
Posted on 23 Feb 2015
Print article
The United States (US) Food and Drug Administration (FDA) has issued a Class 1 Recall notice for a large number of GE Healthcare (Chalfont St. Giles, Buckinghamshire, UK) Magnetic Resonance Imaging (MRI) systems. All MRI systems using superconducting magnets, manufactured from 1985 to the present are affected. The notice was dated February 18th, 2015.

Already on January 6, 2015, GE Healthcare had sent an "Urgent Medical Device Correction" letter to hospital administrators/risk managers, radiology department managers, and radiologists, in the US and around the world. The letter described the safety issue, safety instructions, affected product details, product correction and contact information. The letter was necessary when GE Healthcare discovered that the Magnet Rundown Units (MRU) on its imaging devices may not be properly connected.

To correct the issue, GE Healthcare has instructed customers to perform preventative measures to confirm that the MRU is connected to the magnet. The measures consist of a four-step test on the MRU. The steps are explained on the GE Healthcare and FDA websites.

According the FDA recall notice, “In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries.”

Related Links:

GE Healthcare
US Food and Drug Administration (FDA)


New
Ultrasound Needle Guide
Ultra-Pro 3
40/80-Slice CT System
uCT 528
New
Ultrasound-Guided Biopsy & Visualization Tools
Endoscopic Ultrasound (EUS) Guided Devices
New
Prostate Cancer MRI Analysis Tool
DynaCAD Urology

Print article

Channels

MRI

view channel
Image: The AI tool can help interpret and assess how well treatments are working for MS patients (Photo courtesy of Shutterstock)

AI Tool Tracks Effectiveness of Multiple Sclerosis Treatments Using Brain MRI Scans

Multiple sclerosis (MS) is a condition in which the immune system attacks the brain and spinal cord, leading to impairments in movement, sensation, and cognition. Magnetic Resonance Imaging (MRI) markers... Read more

Imaging IT

view channel
Image: The new Medical Imaging Suite makes healthcare imaging data more accessible, interoperable and useful (Photo courtesy of Google Cloud)

New Google Cloud Medical Imaging Suite Makes Imaging Healthcare Data More Accessible

Medical imaging is a critical tool used to diagnose patients, and there are billions of medical images scanned globally each year. Imaging data accounts for about 90% of all healthcare data1 and, until... Read more